# Script to log and readout PTC data through ethernet connection.
# Tamar Ervin and Jake Spisak, February 2019
# Sanah Bhimani, May 2022
import argparse
import random
import socket
import struct
import time
from ocs import ocs_agent, site_config
from ocs.ocs_twisted import TimeoutLock
STX = '\x02'
ADDR = '\x10'
CMD = '\x80'
CR = '\x0D'
DATA_WRITE = '\x61'
DATA_READ = '\x63'
ESC = '\x07'
ESC_STX = '\x30'
ESC_CR = '\x31'
ESC_ESC = '\x32'
[docs]
class PTC:
def __init__(self, ip_address, port=502, timeout=10, fake_errors=False):
self.ip_address = ip_address
self.port = int(port)
self.fake_errors = fake_errors
self.model = None
self.serial = None
self.software_revision = None
self.comm = socket.socket(socket.AF_INET, socket.SOCK_STREAM)
self.comm.connect((self.ip_address, self.port)) # connects to the PTC
self.comm.settimeout(timeout)
[docs]
def get_data(self):
"""
Gets the raw data from the ptc and returns it in a usable format.
"""
self.comm.sendall(self.buildRegistersQuery())
data = self.comm.recv(1024)
data_flag, brd = self.breakdownReplyData(data)
return data_flag, brd
@staticmethod
def buildRegistersQuery():
query = bytes([0x09, 0x99, # Message ID
0x00, 0x00, # Unused
0x00, 0x06, # Message size in bytes
0x01, # Slave Address
0x04, # Function Code
0x00, 0x01, # The starting Register Number
0x00, 0x35]) # How many to read
return query
[docs]
def power(self, state):
"""Turn the PTC on or off.
Parameters
----------
state : str
Desired power state of the PTC, either 'on', or 'off'.
"""
command = [0x09, 0x99, # Message ID
0x00, 0x00, # Unused
0x00, 0x06, # Message size in bytes
0x01, # Slave Address
0x06, # Function Code
0x00, 0x01] # Register Number
if state.lower() == 'on':
command.extend([0x00, 0x01])
elif state.lower() == 'off':
command.extend([0x00, 0xff])
else:
raise ValueError(f"Invalid state: {state}")
self.comm.sendall(bytes(command))
self.comm.recv(1024) # Discard the echoed command
[docs]
def breakdownReplyData(self, rawdata):
"""Take in raw ptc data, and return a dictionary.
The dictionary keys are the data labels, the dictionary values are the
data in floats or ints.
Returns
-------
data_flag : bool
False if data is valid, True if output could not be interpretted.
data : dict
Data dictionary already formatted for passing to OCS Feed.
"""
# Associations between keys and their location in rawData
keyloc = {"Operating_State": [9, 10],
"Compressor_State": [11, 12],
"Warning_State": [15, 16, 13, 14],
"Alarm_State": [19, 20, 17, 18],
"Coolant_In_Temp": [23, 24, 21, 22],
"Coolant_Out_Temp": [27, 28, 25, 26],
"Oil_Temp": [31, 32, 29, 30],
"Helium_Temp": [35, 36, 33, 34],
"Low_Pressure": [39, 40, 37, 38],
"Low_Pressure_Average": [43, 44, 41, 42],
"High_Pressure": [47, 48, 45, 46],
"High_Pressure_Average": [51, 52, 49, 50],
"Delta_Pressure_Average": [55, 56, 53, 54],
"Motor_Current": [59, 60, 57, 58],
"Hours_of_Operation": [63, 64, 61, 62],
"Pressure_Unit": [65, 66],
"Temperature_Unit": [67, 68],
"Serial_Number": [69, 70],
"Model": [71, 72],
"Software_Revision": [73, 74]}
# Iterate through all keys and return the data in a usable format.
# If there is an error in the string format, print the
# error to logs, return an empty dictionary, and flag the data as bad
data = {}
# If fake_errors=True, then randomly output the string 'FAKE ERROR'
# instead of the actual data 50% of the time
if self.fake_errors:
if random.random() < 0.5:
rawdata = "FAKE ERROR"
try:
for key in keyloc.keys():
locs = keyloc[key]
wkrBytes = bytes([rawdata[loc] for loc in locs])
# four different data formats to unpack
# Big endian unsigned integer 16 bits
if key in [
"Operating_State",
"Compressor_State",
"Pressure_Unit",
"Temperature_Unit",
"Serial_Number",
]:
state = struct.unpack(">H", wkrBytes)[0]
# Serial number is an attribute, not publishable data
if key == "Serial_Number":
self.serial = state
else:
data[key] = state
# 32bit signed integer which is actually stored as a
# 32bit IEEE float (silly)
elif key in ["Warning_State", "Alarm_State"]:
state = int(struct.unpack(">f", wkrBytes)[0])
data[key] = state
# 2 x 8-bit lookup tables.
elif key in ["Model"]:
model_major = struct.unpack(
">B", bytes([rawdata[locs[0]]]))[0]
model_minor = struct.unpack(
">B", bytes([rawdata[locs[1]]]))[0]
# Model is an attribute, not publishable data
self.model = str(model_major) + "_" + str(model_minor)
elif key in ["Software_Revision"]:
version_major = struct.unpack(
">B", bytes([rawdata[locs[0]]]))[0]
version_minor = struct.unpack(
">B", bytes([rawdata[locs[1]]]))[0]
self.software_revision = str(version_major) + "." + str(version_minor)
# 32 bit Big endian IEEE floating point
else:
data[key] = struct.unpack(">f", wkrBytes)[0]
data_flag = False
except BaseException:
data_flag = True
print("Compressor output could not be converted to numbers."
f"Skipping this data block. Bad output string is {rawdata}")
return data_flag, data
def __del__(self):
"""
If the PTC class instance is destroyed, close the connection to the
ptc.
"""
self.comm.close()
[docs]
class PTCAgent:
"""Agent to connect to a single cryomech compressor.
Parameters:
port (int): TCP port to connect to.
ip_address (str): IP Address for the compressor.
f_sample (float, optional): Data acquisiton rate, defaults to 2.5 Hz.
fake_errors (bool, optional): Generates fake errors in the string
output 50% of the time.
"""
def __init__(self, agent, port, ip_address, f_sample=2.5,
fake_errors=False):
self.agent = agent
self.log = agent.log
self.lock = TimeoutLock()
self.ip_address = ip_address
self.fake_errors = fake_errors
self.port = port
self.module = None
self.f_sample = f_sample
self.initialized = False
self.take_data = False
# Registers data feeds
agg_params = {
'frame_length': 60,
}
self.agent.register_feed('ptc_status',
record=True,
agg_params=agg_params,
buffer_time=1)
[docs]
@ocs_agent.param('auto_acquire', default=False, type=bool)
def init(self, session, params=None):
"""init(auto_acquire=False)
**Task** - Initializes the connection to the PTC.
Parameters:
auto_acquire (bool): Automatically start acq process after
initialization if True. Defaults to False.
"""
if self.initialized:
return True, "Already Initialized"
with self.lock.acquire_timeout(0, job='init') as acquired:
if not acquired:
self.log.warn("Could not start init because "
"{} is already running".format(self.lock.job))
return False, "Could not acquire lock."
# Establish connection to ptc
self.ptc = PTC(self.ip_address, port=self.port,
fake_errors=self.fake_errors)
# Test connection and display identifying info
self.ptc.get_data()
print("PTC Model:", self.ptc.model)
print("PTC Serial Number:", self.ptc.serial)
print("Software Revision is:", self.ptc.software_revision)
self.initialized = True
# Start data acquisition if requested
if params['auto_acquire']:
resp = self.agent.start('acq', params={})
self.log.info(f'Response from acq.start(): {resp[1]}')
return True, "PTC agent initialized"
[docs]
@ocs_agent.param('state', type=str, choices=['off', 'on'])
def power_ptc(self, session, params=None):
"""power_ptc(state=None)
**Task** - Remotely turn the PTC on or off.
Parameters
----------
state : str
Desired power state of the PTC, either 'on', or 'off'.
"""
with self.lock.acquire_timeout(3, job='power_ptc') as acquired:
if not acquired:
self.log.warn("Could not start task because {} is already "
"running".format(self.lock.job))
return False, "Could not acquire lock."
self.ptc.power(params['state'])
return True, "PTC powered {}".format(params['state'])
[docs]
@ocs_agent.param('test_mode', default=False, type=bool)
def acq(self, session, params):
"""acq()
**Process** - Starts acqusition of data from the PTC.
Parameters:
test_mode (bool, optional): Run the Process loop only once.
This is meant only for testing. Default is False.
"""
with self.lock.acquire_timeout(0, job='acq') as acquired:
if not acquired:
self.log.warn("Could not start acq because {} is already"
"running".format(self.lock.job))
return False, "Could not acquire lock."
last_release = time.time()
self.take_data = True
while self.take_data:
# Relinquish sampling lock occasionally
if time.time() - last_release > 1.:
last_release = time.time()
if not self.lock.release_and_acquire(timeout=10):
self.log.warn(f"Failed to re-acquire sampling lock, "
f"currently held by {self.lock.job}.")
continue
# Publish data, waiting 1/f_sample seconds in between calls.
pub_data = {'timestamp': time.time(),
'block_name': 'ptc_status'}
data_flag, data = self.ptc.get_data()
pub_data['data'] = data
# If there is an error in compressor output (data_flag = True),
# do not publish
if not data_flag:
self.agent.publish_to_feed('ptc_status', pub_data)
time.sleep(1. / self.f_sample)
if params['test_mode']:
break
self.agent.feeds["ptc_status"].flush_buffer()
return True, 'Acquisition exited cleanly.'
def _stop_acq(self, session, params):
"""Stops acqusition of data from the PTC."""
if self.take_data:
self.take_data = False
return True, 'requested to stop taking data.'
else:
return False, 'acq is not currently running'
def make_parser(parser=None):
"""Build the argument parser for the Agent. Allows sphinx to automatically
build documentation based on this function.
"""
if parser is None:
parser = argparse.ArgumentParser()
# Add options specific to this agent.
pgroup = parser.add_argument_group('Agent Options')
pgroup.add_argument('--ip-address')
pgroup.add_argument('--port', default=502)
pgroup.add_argument('--serial-number')
pgroup.add_argument('--mode', choices=['init', 'acq'])
pgroup.add_argument('--fake-errors', default=False,
help="If True, randomly output 'FAKE ERROR' instead of "
"data half of the time.")
return parser
def main(args=None):
parser = make_parser()
args = site_config.parse_args(agent_class='CryomechCPAAgent',
parser=parser,
args=args)
print('I am in charge of device with serial number: %s' % args.serial_number)
# Automatically acquire data if requested (default)
init_params = False
if args.mode == 'init':
init_params = {'auto_acquire': False}
elif args.mode == 'acq':
init_params = {'auto_acquire': True}
# Call launcher function (initiates connection to appropriate
# WAMP hub and realm).
agent, runner = ocs_agent.init_site_agent(args)
# create agent instance and run log creation
ptc = PTCAgent(agent, args.port, args.ip_address,
fake_errors=args.fake_errors)
agent.register_task('init', ptc.init, startup=init_params)
agent.register_process('acq', ptc.acq, ptc._stop_acq)
agent.register_task('power_ptc', ptc.power_ptc)
runner.run(agent, auto_reconnect=True)
if __name__ == '__main__':