Source code for socs.agents.lakeshore336.agent

#!/usr/bin/env python3

# Author: zatkins, zhuber
# Acknowledgments: LS372 agent -- bkoopman, mhasselfield, jlashner

import argparse
import time

import numpy as np
from ocs import ocs_agent, site_config
from ocs.ocs_twisted import TimeoutLock

from socs.Lakeshore.Lakeshore336 import LS336


[docs] class LS336_Agent: """Agent to connect to a single Lakeshore 336 device. Supports channels 'A','B','C', and 'D' for Lakeshore 336s that do not have the extra Lakeshore 3062 scanner installed. Also has channels 'D2','D3','D4', and 'D5' for 336s that have the extra scanner. Currently only supports heaters '1' and '2'. Parameters ---------- sn: str Serial number of the LS336 ip: str IP Address for the 336 device f_sample: float, optional (default 0.1) The frequency of sampling for acquiring data (in Hz) threshold: float, optional (default 0.1) The max difference (in K) between the setpoint and current temperature that will be considered stable window: int, optional (default 900) The amount of time (in s) over which the difference between the setpoint and the current temperature must not exceed threshold while checking for temperature stability. Attributes ---------- sn: str Serial number of the LS336 ip: str IP Address for the 336 device module: LS336 object Driver object module.channels: dict The available channels in the LS336 object module.heaters: dict The available heaters in the LS336 object f_sample: float The frequency of sampling for acquiring data (in Hz) t_sample: float The time between each sample (inverse of f_sample - 0.01) threshold: float The max difference (in K) between the setpoint and current temperature that will be considered stable window: int The amount of time (in s) over which the difference between the setpoint and the current temperature must not exceed threshold while checking for temperature stability. _recent_temps: numpy array, protected Array of recent temperatures for checking temperature stability _static_setpoint: float, protected The final setpoint value to avoid issues when the setpoint is ramping to a new value. Used in checking temperature stability """ def __init__(self, agent, sn, ip, f_sample=0.1, threshold=0.1, window=900): self.agent = agent self.sn = sn self.ip = ip self.f_sample = f_sample self.t_sample = 1 / self.f_sample - 0.01 assert self.t_sample < 7200, \ "acq sampling freq must be such that t_sample is less than 2 hours" self._lock = TimeoutLock() self.log = agent.log self.initialized = False self.take_data = False self.module = None # for stability checking self.threshold = threshold self.window = window self._recent_temps = None self._static_setpoint = None agg_params = {'frame_length': 10 * 60} # sec # combined feed for thermometry and control data self.agent.register_feed( 'temperatures', record=True, agg_params=agg_params, buffer_time=1 )
[docs] @ocs_agent.param('auto_acquire', default=False, type=bool) def init_lakeshore(self, session, params=None): """init_lakeshore(auto_acquire=False) **Task** - Perform first time setup of the Lakeshore 336 communication Parameters: auto_acquire (bool, optional): Default is False. Starts data acquisition after initialization if True. """ if params is None: params = {} # test if this agent is already running if self.initialized: self.log.info('Lakeshore already initialized, returning...') return True, 'Already initialized' # initialize lakeshore with self._lock.acquire_timeout(job='init', timeout=0) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get lakeshore self.module = LS336(self.ip) session.add_message( f'Lakeshore initialized with ID: {self.module.id}') self.initialized = True # start data acq if passed if params.get('auto_acquire', False): self.agent.start('acq') return True, 'Lakeshore module initialized'
[docs] @ocs_agent.param('f_sample', default=0.1, type=float) def acq(self, session, params=None): """acq(f_sample=0.1) **Process** - Begins recording data from thermometers and heaters. Parameters: f_sample (float, optional): Default is 0.1. Sets the sampling rate in Hz. Notes: The most recent data collected is stored in session.data in the structure: >>> response.session['data'] {"ls336_fields": {"timestamp": 1921920543, "block_name": "temperatures" "data": {"Channel_A_T": (some value) "Channel_A_V": (some value) "Channel_B_T": (some value) "Channel_B_V": (some value) "Channel_C_T": (some value) "Channel_C_V": (some value) "Channel_D_T": (some value) "Channel_D_V": (some value) } } } """ if params is None: params = {} # get sampling frequency f_sample = params.get('f_sample') if f_sample is None: t_sample = self.t_sample else: t_sample = 1 / f_sample - 0.01 self.t_sample = t_sample # acquire lock and start Process with self._lock.acquire_timeout(job='acq', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # initialize recent temps array # shape is N_points x N_channels # N_points is 2 hour / t_sample rounded up # N_channels is 8 if the extra scanner is installed, 4 otherwise # t_sample can't be more than 2 hours N_channels = len(self.module.channels) self._recent_temps = np.full( (int(np.ceil(7200 / self.t_sample)), N_channels), -1.0) print(self._recent_temps.size) # acquire data from Lakeshore self.take_data = True while self.take_data: # get thermometry data current_time = time.time() temperatures_message = { 'timestamp': current_time, 'block_name': 'temperatures', 'data': {} } temps = self.module.get_kelvin('0') # array of 4 (or 8) floats voltages = self.module.get_sensor('0') # array of 4/8 floats for i, channel in enumerate(self.module.channels.values()): channel_str = channel.input_name.replace(' ', '_') temperatures_message['data'][channel_str + '_T'] = temps[i] temperatures_message['data'][channel_str + '_V'] = voltages[i] # append to recent temps array for temp stability check self._recent_temps = np.roll(self._recent_temps, 1, axis=0) self._recent_temps[0] = temps # publish to feed self.agent.publish_to_feed( 'temperatures', temperatures_message) # For session.data - named to avoid conflicting with LS372 # if in use at same time. session.data['ls336_fields'] = temperatures_message # get heater data heaters_message = { 'timestamp': current_time, 'block_name': 'heaters', 'data': {} } for i, heater in enumerate(self.module.heaters.values()): heater_str = heater.output_name.replace(' ', '_') heaters_message['data'][ heater_str + '_Percent'] = heater.get_heater_percent() heaters_message['data'][ heater_str + '_Range'] = heater.get_heater_range() heaters_message['data'][ heater_str + '_Max_Current'] = heater.get_max_current() heaters_message['data'][ heater_str + '_Setpoint'] = heater.get_setpoint() # publish to feed self.agent.publish_to_feed('temperatures', heaters_message) # finish sample self.log.debug( f'Sleeping for {np.round(self.t_sample)} seconds...') # release and reacquire lock between data acquisition self._lock.release() time.sleep(t_sample) if not self._lock.acquire(timeout=10, job='acq'): print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not re-acquire lock' return True, 'Acquisition exited cleanly'
[docs] @ocs_agent.param('_') def stop_acq(self, session, params=None): """stop_acq() **Task** - Stops acq process. """ if params is None: params = {} if self.take_data: self.take_data = False return True, 'Requested to stop taking data' else: return False, 'acq is not currently running'
[docs] @ocs_agent.param('range', type=str, choices=['off', 'low', 'medium', 'high']) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_heater_range(self, session, params): """set_heater_range(range=None,heater='2') **Task** - Adjusts the heater range for servoing the load. Parameters: range (str): Sets the range of the chosen heater. Must be one of 'off', 'low', 'medium', and 'high'. heater (str, optional): default '2'. Chooses which heater's range to change. Must be '1' or '2'. Notes: The range setting has no effect if an output is in the Off mode, and it does not apply to an output in Monitor Out mode. An output in Monitor Out mode is always on. """ with self._lock.acquire_timeout(job='set_heater_range', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set range current_range = heater.get_heater_range() if params['range'] == current_range: print( 'Current heater range matches commanded value. ' 'Proceeding unchanged') else: heater.set_heater_range(params['range']) session.add_message( f"Set {heater.output_name} range to {params['range']}") return True, f"Set {heater.output_name} range to {params['range']}"
[docs] @ocs_agent.param('P', type=float, check=lambda x: 0.1 <= x <= 1000) @ocs_agent.param('I', type=float, check=lambda x: 0.1 <= x <= 1000) @ocs_agent.param('D', type=float, check=lambda x: 0 <= x <= 200) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_pid(self, session, params): """set_pid(P=None,I=None,D=None,heater='2') **Task** - Set the PID parameters for servoing the load. Parameters: P (float): Proportional term for PID loop (must be between 0.1 and 1000) I (float): Integral term for PID loop (must be between 0.1 and 1000) D (float): Derivative term for PID loop (must be between 0 and 200) heater (str, optional): Selects the heater on which to change the PID settings. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_pid', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set pid current_p, current_i, current_d = heater.get_pid() if (params['P'] == current_p and params['I'] == current_i and params['D'] == current_d): print('Current heater PID matches commanded value. ' 'Proceeding unchanged') else: heater.set_pid(params['P'], params['I'], params['D']) session.add_message( f"Set {heater.output_name} PID to {params['P']}, " f"{params['I']}, {params['D']}") return True, (f"Set {heater.output_name} PID to {params['P']}, " f" {params['I']}, {params['D']}")
[docs] @ocs_agent.param('mode', type=str, choices=['off', 'closed loop', 'zone', 'open loop']) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_mode(self, session, params): """set_mode(mode=None,heater='2') **Task** - Sets the output mode of the heater. Parameters: mode (str): Selects the output mode for the heater. Accepts four options: 'off', 'closed loop', 'zone', and 'open loop'. for restrictions based on the selected heater. heater (str, optional): Default '2'. Selects the heater on which to change the mode. Must be '1' or '2'. Notes: Does not support the options 'monitor out' and 'warm up', which only work for the unsupported analog outputs (heaters 3 and 4). """ with self._lock.acquire_timeout(job='set_mode', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set mode current_mode = heater.get_mode() if params['mode'] == current_mode: print( 'Current heater mode matches commanded value. ' 'Proceeding unchanged') else: heater.set_mode(params['mode']) session.add_message( f"Set {heater.output_name} mode to {params['mode']}") return True, f"Set {heater.output_name} mode to {params['mode']}"
[docs] @ocs_agent.param('resistance', type=float) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_heater_resistance(self, session, params): """set_heater_resistance(resistance=None,heater='2') **Task** - Sets the heater resistance and resistance setting of the heater. The associated 'get' function in the Heater class is get_heater_resistance_setting(). Parameters: resistance (float): The actual resistance of the load heater (str, optional): Default '2'. Selects the heater on which to set the resistance. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_heater_resistance', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set heater resistance _ = heater.get_heater_resistance_setting() if params['resistance'] == heater.resistance: print( 'Current heater resistance matches commanded value. ' 'Proceeding unchanged') else: heater.set_heater_resistance(params['resistance']) session.add_message( f"Set {heater.output_name} resistance to " f"{params['resistance']}") return True, (f"Set {heater.output_name} resistance to " f"{params['resistance']}")
[docs] @ocs_agent.param('current', type=float, check=lambda x: 0.0 <= x <= 2.0) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_max_current(self, session, params): """set_max_current(current=None,heater='2') **Task** - Sets the maximum current that can pass through a heater. Parameters: current (float): The desired max current. Must be between 0 and 2 A. heater (str, optional): Default '2'. Selects the heater on which to set the max current. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_max_current', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set max current current_max_current = heater.get_max_current() if params['current'] == current_max_current: print( 'Current max current matches commanded value. ' 'Proceeding unchanged') else: heater.set_max_current(params['current']) session.add_message( f"Set {heater.output_name} max current to {params['current']}") return True, (f"Set {heater.output_name} max current to " f"{params['current']}")
[docs] @ocs_agent.param('percent', type=float) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_manual_out(self, session, params): """set_manual_out(percent=None,heater='2') **Task** - Sets the manual output of the heater as a percentage of the full current or power depending on which display the heater is set to use. Parameters: percent (float): Percent of full current or power to set on the heater. Must have 2 or fewer decimal places. heater (str, optional): Default '2'. Selects the heater on which to set the manual output. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_manual_out', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set manual out current_manual_out = heater.get_manual_out() if params['percent'] == current_manual_out: print('Current manual out matches commanded value. ' 'Proceeding unchanged') else: heater.set_manual_out(params['percent']) session.add_message( f"Set {heater.output_name} manual out to {params['percent']}") return True, (f"Set {heater.output_name} manual out to " f"{params['percent']}")
[docs] @ocs_agent.param('input', type=str, choices=['A', 'B', 'C', 'D', 'D1', 'D2', 'D3', 'D4', 'D5']) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_input_channel(self, session, params): """set_input_channel(input=None,heater='2') **Task** - Sets the input channel of the heater control loop. Parameters: input (str): The name of the heater to use as the input channel. Must be one of 'none','A','B','C', or 'D'. Can also be 'D2','D3','D4', or 'D5' if the extra Lakeshore 3062 Scanner is installed in your LS336. heater (str, optional): Default '2'. Selects the heater for which to set the input channel. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_input_channel', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # D1 is the same as D if params['input'] == 'D1': params['input'] = 'D' # set input channel current_input_channel = heater.get_input_channel() if params['input'] == current_input_channel: print( 'Current input channel matches commanded value. ' 'Proceeding unchanged') else: heater.set_input_channel(params['input']) session.add_message( f"Set {heater.output_name} input channel to {params['input']}") return True, (f"Set {heater.output_name} input channel to " f"{params['input']}")
[docs] @ocs_agent.param('setpoint', type=float) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def set_setpoint(self, session, params): """set_setpoint(setpoint=None,heater='2') **Task** - Sets the setpoint of the heater control loop, after first turning ramp off. May be a limit to how high the setpoint can go based on your system parameters. Parameters: setpoint (float): The setpoint for the control loop. Units depend on the preferred sensor units (Kelvin, Celsius, or Sensor). heater (str, optional): Default '2'. Selects the heater for which to set the input channel. Must be '1' or '2'. """ with self._lock.acquire_timeout(job='set_setpoint', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # set setpoint current_setpoint = heater.get_setpoint() if params['setpoint'] == current_setpoint: print('Current setpoint matches commanded value. ' 'Proceeding unchanged') else: heater.set_ramp_on_off('off') heater.set_setpoint(params['setpoint']) # static setpoint used in temp stability check # to avoid ramping bug self._static_setpoint = params['setpoint'] session.add_message( f"Turned ramp off and set {heater.output_name} setpoint to " f"{params['setpoint']}") return True, (f"Turned ramp off and set {heater.output_name} " f"setpoint to {params['setpoint']}")
[docs] @ocs_agent.param('T_limit', type=int) @ocs_agent.param('channel', type=str, default='A', choices=['A', 'B', 'C', 'D', 'D2', 'D3', 'D4', 'D5']) def set_T_limit(self, session, params): """set_T_limit(T_limit=None,channel='A') **Task** - Sets the temperature limit above which the control output assigned to the selected channel shut off. Parameters: T_limit (int): The temperature limit in Kelvin. Note that a limit of 0 K turns off this feature for the given channel. channel (str, optional): Default 'A'. Selects which channel to use for controlling the temperature. Options are 'A','B','C', and 'D'. Can also be 'D2','D3','D4', or 'D5' if the extra Lakeshore 3062 Scanner is installed in your LS336. """ with self._lock.acquire_timeout(job='set_T_limit', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get channel channel_key = params.get('channel', 'A') # default to input A channel = self.module.channels[channel_key] # set T limit current_limit = channel.get_T_limit() if params['T_limit'] == current_limit: print('Current T limit matches commanded value. ' 'Proceeding unchanged') else: channel.set_T_limit(params['T_limit']) session.add_message( f"Set {channel.input_name} T limit to {params['T_limit']}") return True, f"Set {channel.input_name} T limit to {params['T_limit']}"
[docs] @ocs_agent.param('temperature', type=float) @ocs_agent.param('ramp', default=0.1, type=float) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) @ocs_agent.param('transport', default=False, type=bool) @ocs_agent.param('transport_offset', default=0, type=float, check=lambda x: x >= 0.0) def servo_to_temperature(self, session, params): """servo_to_temperature(temperature=None,ramp=0.1,heater='2',\ transport=False,transport_offset=0) **Task** - A wrapper for setting the heater setpoint. Performs sanity checks on heater configuration before publishing setpoint: 1. checks control mode of heater (closed loop) 2. checks units of input channel (kelvin) 3. resets setpoint to current temperature with ramp off 4. sets ramp on to specified rate 5. checks setpoint does not exceed input channel T_limit 6. sets setpoint to commanded value Note that this function does NOT turn on the heater if it is off. You must use set_heater_range() to pick a range first. Parameters: temperature (float): The new setpoint in Kelvin. Make sure there is is a control input set to the heater and its units are Kelvin. ramp (float, optional): Default 0.1. The rate for how quickly the setpoint ramps to new value. Units of K/min. heater (str, optional): Default '2'. The heater to use for servoing. Must be '1' or '2'. transport (bool, optional): Default False. See Notes for description. transport_offset (float, optional): Default 0. In Kelvin. See Notes. Notes: If param 'transport' is provided and True, the control loop restarts when the setpoint is first reached. This is useful for loads with long cooling times or time constant to help minimize over/undershoot. If param 'transport' is provided and True, and 'transport_offset' is provided and positive, and the setpoint is higher than the current temperature, then the control loop will restart when the setpoint - transport_offset is first reached. This is useful to avoid a "false positive" temperature stability check too shortly after transport completes. """ # get sampling frequency t_sample = self.t_sample with self._lock.acquire_timeout(job='servo_to_temperature', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # get current setpoint current_setpoint = heater.get_setpoint() # check in correct control mode if heater.get_mode() != 'closed loop': session.add_message( 'Changing control to closed loop mode for servo.') heater.set_mode("closed loop") # check in correct units channel = heater.get_input_channel() if channel == 'none': return False, (f'{heater.output_name} does not have an ' f'input channel assigned') if self.module.channels[channel].get_units() != 'kelvin': session.add_message( 'Setting preferred units to kelvin on ' 'heater control input.') self.module.channels[channel].set_units('kelvin') # restart setpoint at current temperature current_temp = np.round(float(self.module.get_kelvin(channel)), 4) session.add_message( f'Turning ramp off and setting setpoint to current ' f'temperature {current_temp}') heater.set_ramp_on_off('off') heater.set_setpoint(current_temp) # reset ramp settings ramp = params.get('ramp', 0.1) session.add_message( f'Turning ramp on and setting rate to {ramp}K/min') heater.set_ramp_on_off('on') heater.set_ramp_rate(ramp) # make sure not exceeding channel T limit T_limit = self.module.channels[channel].get_T_limit() if T_limit <= params['temperature']: return False, (f"{heater.output_name} control channel " f"{channel} T limit of {T_limit}K is lower " f"than setpoint of {params['temperature']}") # set setpoint if params['temperature'] == current_setpoint: print('Current setpoint matches commanded value. ' 'Proceeding unchanged') else: session.add_message( f"Setting {heater.output_name} setpoint to " f"{params['temperature']}") heater.set_setpoint(params['temperature']) # static setpoint used in temp stability check # to avoid pulling the ramping setpoint self._static_setpoint = params['temperature'] # if transport, restart control loop when setpoint # first crossed if params.get('transport', False): current_range = heater.get_heater_range() starting_sign = np.sign( params['temperature'] - current_temp) transporting = True # if we are raising temp, allow possibility of # stopping transport at a cooler temp T_offset = 0 if starting_sign > 0: T_offset = params.get('transport_offset', 0) if T_offset < 0: return False, ('Transport offset temperature ' 'cannot be negative') while transporting: current_temp = np.round( self.module.get_kelvin(channel), 4) # check when this flips current_sign = np.sign( params['temperature'] - T_offset - current_temp) # release and reacquire lock between data acquisition self._lock.release() time.sleep(t_sample) if not self._lock.acquire(timeout=10, job='servo_to_temperature'): print( f"Lock could not be acquired because it is " f"held by {self._lock.job}") return False, 'Could not re-acquire lock' if current_sign != starting_sign: transporting = False # update flag # cycle control loop session.add_message( 'Transport complete, restarting control ' 'loop at provided setpoint') heater.set_heater_range('off') # necessary 1 s for prev command to register # in ls336 firmware for some reason time.sleep(1) heater.set_heater_range(current_range) return True, (f"Set {heater.output_name} setpoint to " f"{params['temperature']}")
[docs] @ocs_agent.param('threshold', default=0.1, type=float) @ocs_agent.param('window', default=900, type=int) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def check_temperature_stability(self, session, params): """check_temperature_stability(threshold=0.1,window=900,heater='2') **Task** - Assesses whether the load is stable around the setpoint to within some threshold. Parameters: threshold (float, optional): Default 0.1. See notes. window (int, optional): Default 900. See notes. heater (str, optional): Default '2'. Selects the heater for which to set the input channel. Must be '1' or '2'. Notes ----- Param 'threshold' sets the upper bound on the absolute temperature difference between the setpoint and any temperature from the input channel in the last 'window' seconds. Param 'window' sets the lookback time into the most recent temperature data, in seconds. Note that this function grabs the most recent data in one window-length of time; it does not take new data. If you want to use the result of this task for making logical decisions in a client (e.g. waiting longer before starting a process if the temperature is not yet stable), use the session['success'] key. It will be True if the temperature is stable and False if not. Example: >>> response = ls336.check_temperature_stability() >>> response.session['success'] True """ # get threshold threshold = params.get('threshold') if threshold is None: threshold = self.threshold # get window window = params.get('window') if window is None: window = self.window num_idxs = int(np.ceil(window / self.t_sample)) with self._lock.acquire_timeout(job='check_temperature_stability', timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # get channel channel = heater.get_input_channel() channel_num = self.module.channels[channel].num # get current temp current_temp = np.round(self.module.get_kelvin(channel), 4) # check if recent temps and current temps are within threshold _recent_temps = self._recent_temps[:num_idxs, channel_num - 1] _recent_temps = np.concatenate( (np.array([current_temp]), _recent_temps)) # get static setpoint if None if self._static_setpoint is None: self._static_setpoint = heater.get_setpoint() # avoids checking against the ramping setpoint, # i.e. want to compare to commanded setpoint not mid-ramp setpoint setpoint = self._static_setpoint session.add_message( f'Maximum absolute difference in recent temps is ' f'{np.max(np.abs(_recent_temps - setpoint))}K') if np.all(np.abs(_recent_temps - setpoint) < threshold): session.add_message( f'Recent temps are within {threshold}K of setpoint') return True, (f'Servo temperature is stable within ' f'{threshold}K of setpoint') return False, (f'Servo temperature is not stable within ' f'{threshold}K of setpoint')
[docs] @ocs_agent.param('attribute', type=str) @ocs_agent.param('channel', type=str, default='A', choices=['A', 'B', 'C', 'D', 'D1', 'D2', 'D3', 'D4', 'D5']) def get_channel_attribute(self, session, params): """get_channel_attribute(attribute=None,channel='A') **Task** - Gets an arbitrary channel attribute and stores it in the session.data dict. Attribute must be the name of a method in the namespace of the Lakeshore336 Channel class, with a leading "get\\_" removed (see example). Parameters: attribute (str): The name of the channel attribute to get. See the Lakeshore 336 Channel class API for all options. channel (str, optional): Default 'A'. Selects which channel for which to get the attribute. Options are 'A','B','C', and 'D'. Can also be 'D2','D3','D4', or 'D5' if the extra Lakeshore 3062 Scanner is installed in your LS336. Example: >>> ls.get_channel_attribute(attribute = 'T_limit').session['data'] {'T_limit': 30.0} """ with self._lock.acquire_timeout(job=f"get_{params['attribute']}", timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get channel channel_key = params.get('channel', 'A') # default to input A channel = self.module.channels[channel_key] # check that attribute is a valid channel method if getattr(channel, f"get_{params['attribute']}", False) is not False: query = getattr(channel, f"get_{params['attribute']}") # get attribute resp = query() session.data[params['attribute']] = resp return True, (f"Retrieved {channel.input_name} {params['attribute']}, value is {resp}")
[docs] @ocs_agent.param('attribute', type=str) @ocs_agent.param('heater', default='2', type=str, choices=['1', '2']) def get_heater_attribute(self, session, params): """get_heater_attribute(attribute=None,heater='2') **Task** - Gets an arbitrary heater attribute and stores it in the session.data dict. Attribute must be the name of a method in the namespace of the Lakeshore336 Heater class, with a leading "get\\_" removed (see example). Parameters: attribute (str): The name of the channel attribute to get. See the Lakeshore 336 Heater class API for all options. heater (str, optional): Default '2'. Selects the heater for which to get the heater attribute. Must be '1' or '2'. Examples -------- >>> ls.get_heater_attribute(attribute = 'heater_range').session['data'] {'heater_range': 'off'} """ with self._lock.acquire_timeout(job=f"get_{params['attribute']}", timeout=3) as acquired: if not acquired: print( f"Lock could not be acquired because it is held by " f"{self._lock.job}") return False, 'Could not acquire lock' # get heater heater_key = params.get('heater', '2') # default to 50W output heater = self.module.heaters[heater_key] # check that attribute is a valid heater method if getattr(heater, f"get_{params['attribute']}", False) is not False: query = getattr(heater, f"get_{params['attribute']}") # get attribute resp = query() session.data[params['attribute']] = resp return True, f"Retrieved {heater.output_name} {params['attribute']}, value is {resp}"
def make_parser(parser=None): """Build the argument parser for the Agent. Allows sphinx to automatically build documentation based on this function. """ if parser is None: parser = argparse.ArgumentParser() # Add options specific to this agent. pgroup = parser.add_argument_group('Agent Options') pgroup.add_argument('--serial-number') pgroup.add_argument('--ip-address', type=str, help="IP address for the lakeshore") pgroup.add_argument('--f-sample', type=float, default=0.1, help='The frequency of data acquisition') pgroup.add_argument('--threshold', type=float, default=0.1, help='The upper bound on temperature differences ' 'for stability check') pgroup.add_argument('--window', type=float, default=900., help='The lookback time on temperature differences ' 'for stability check') pgroup.add_argument('--auto-acquire', type=bool, default=True, help='Automatically start data acquisition on startup') return parser def main(args=None): # Create an argument parser parser = make_parser() args = site_config.parse_args(agent_class='Lakeshore336Agent', parser=parser, args=args) # Automatically acquire data if requested init_params = False if args.auto_acquire: init_params = {'auto_acquire': True} print('I am in charge of device with serial ' 'number: %s' % args.serial_number) # Create a session and a runner which communicate over WAMP agent, runner = ocs_agent.init_site_agent(args) # Pass the new agent session to the agent class lake_agent = LS336_Agent(agent, args.serial_number, args.ip_address, args.f_sample, args.threshold, args.window) # Register tasks (name, agent_function) agent.register_task( 'init_lakeshore', lake_agent.init_lakeshore, startup=init_params) agent.register_task('set_heater_range', lake_agent.set_heater_range) agent.register_task('set_heater_resistance', lake_agent.set_heater_resistance) agent.register_task('set_input_channel', lake_agent.set_input_channel) agent.register_task('set_manual_out', lake_agent.set_manual_out) agent.register_task('set_max_current', lake_agent.set_max_current) agent.register_task('set_mode', lake_agent.set_mode) agent.register_task('set_pid', lake_agent.set_pid) agent.register_task('set_T_limit', lake_agent.set_T_limit) agent.register_task('set_setpoint', lake_agent.set_setpoint) agent.register_task('servo_to_temperature', lake_agent.servo_to_temperature) agent.register_task('check_temperature_stability', lake_agent.check_temperature_stability) agent.register_task('get_channel_attribute', lake_agent.get_channel_attribute) agent.register_task('get_heater_attribute', lake_agent.get_heater_attribute) # Register processes (name, agent_start_function, agent_end_function) agent.register_process('acq', lake_agent.acq, lake_agent.stop_acq) # Run the agent runner.run(agent, auto_reconnect=True) if __name__ == '__main__': main()